Color OLED 320×240 Pixels Display
Live Simultaneous Color A-Scan and Thickness Display
Control of Gain,Blanking,Gate,Range,Delay,RF and Rectify Modes
Time-based B-Scan Display
Echo-Echo Mode(Thru Paint & Coatings)
Stores 100,000 Readings and 1000 waveforms
Color OLED 320×240 Pixels Display
Live Simultaneous Color A-Scan and Thickness Display
Control of Gain,Blanking,Gate,Range,Delay,RF and Rectify Modes
Time-based B-Scan Display
Echo-Echo Mode(Thru Paint & Coatings)
Stores 100,000 Readings and 1000 waveforms
TM280 Series Technical Specifications
| Feature | TM280 | TM280D | TM280DL | Data Logger Option |
| Color Display | √ | √ | √ | |
| LiveA-Scan | √ | √ | √ | |
| Time-based B-Scan | √ | √ | √ | |
| Control of Gain, Blanking,Gate, Range, Delay, RFand Rectify Modes | √ | √ | √ | |
| Echo-Echo Mode
(Thru-paint &Coatings) |
× | √ | √ | |
| Data Logger | × | × | √ | √ |
| DataView Software | × | × | √ | √ |
| Instrument Specifications | |
| Display Type | 2.4” color OLED, 320 X 240 pixels, contrast 10,000:1 |
| Operating Principle | Pulse echo with dual element transducers |
| Measuring Range | 0.50mm to 508mm(0.02” to 20.00″), depending on material, probe and surface
condition |
| Measuring Resolution | Selectable 0.01mm, 0.1mm(selectable 0.001″, 0.01″) |
| Units | Inch or Millimeter |
| Rectify Modes | RF+, RF-, HALF+, HALF-, FULL |
| Display Mode | Normal, Minimum / Maximum capture, DIFF/RR%, A-Scan, B-Scan |
| V-Path Correction | Automatic |
| Update Rate | Selectable 4Hz, 8Hz, 16Hz |
| Material Velocity Range | 500 to 9999m/s (0.0197 to 0.3937in/us) |
| Languages | English |
| Alarm Settings | Minimum and Maximum alarms. Range of 0.25 mm to 508 mm (0.010″ to 20.00″). Dynamic waveform color change on alarm |
| Power Requirements | 2 AA size batteries |
| Operating Time | Approximately 40 hours |
| Instrument Shut-off | Selectable ALWAYS ON or AUTO OFF after 5, 10, 20 minutes of inactivity |
| Operating Temperature | -10°C to +50°C (+10°F to +120°F) |
| Size | 153mm X 76mm X 37mm(H X W X D) |
| Weight | 280g including batteries |
| Data Logger Option Features | |
| Capacity | 100,000 readings,1000 waveforms, 400 files |
| File Structure | Grid file |
| Rows X Columns | 21 X 12 |
| Communication Port | USB 2.0 (Full Speed ) port |
| Software | DataView PC software |
| TM280 Probe/Transducer Specifications | |||||||
| Model | PT08 | TC510 | PT12 | ZT12 | PT06 | PT04 | GT12 |
| Type | TM280
Standard |
TM280D/DLStandard | Standard | Cast iron
|
Small tube | Fingertip | High-Temperature |
| Frequency | 5MHz | 5MHz | 5MHz | 2MHz | 7.5MHz | 10MHz | 5MHz |
| Contact Diameter | 11mm | 13.5mm | 13.5mm | 17mm | 8mm | 6mm | 15mm |
| Measurement
Range |
0.8 to
100.0mm |
1.2 to
200.0mm |
1.0 to
200.0mm |
4.0 to
508mm |
0.8 to
30mm |
0.7 to
12mm |
4 to
80mm |
| TemperatureRange | -10 to 70℃ | -10 to 70℃ | -10 to 70℃ | -10 to 70℃ | -10 to 70℃ | -10 to 70℃ | -20 to 300℃ |